MIXED FLOW GAS TESTING PER ASTM B827 & ASTM B845

Mixed flowing gas (MFG) is a type of laboratory environmental testing for products, particularly electronics, to evaluate resistance to corrosion due to gases in the atmosphere.

GET IN TOUCH

What is it?

The purpose of this test is to simulate corrosion phenomenon due to atmospheric exposure.

Mixed Flowing Gas (MFG) test is a laboratory test in which the temperature (°C), relative humidity (%RH), concentration of gaseous pollutants (in parts per billion, ppb or parts per million ppm level), and other critical variables (such as volume exchange rate and airflow rate) are carefully defined, monitored and controlled.

The electronic product is exposed to gases such as chlorine, hydrogen sulfide, sulfide, nitrogen dioxide and sulfur dioxide at levels in the ppb range, in a controlled environmental chamber. Test samples that have been exposed to MFG testing have ranged from bare metal surfaces, to electrical connectors, and to complete assemblies. In regards to noble metal plated connector applications, MFG testing has been widely accepted as a qualification test method to evaluate the performance of these connectors. With regard to electrical and electronic applications, MFG testing has been widely accepted as a qualification test method to evaluate performance in the presence of environmental pollutants. The MFG exposures are generally used in conjunction with procedures which evaluate contact or connector electrical performance, such as measurement of electrical contact resistance before and after MFG exposure.

With regard to noble metal plated connector applications, MFG testing has been widely accepted as a qualification test method for evaluating the performance of connectors.

Matergenics MFG testing chamber attributes:

  •  Multiple-gas atmospheres controllable at the parts per billion (ppb) level
  •  No time limit on exposure
  •  Temperature and humidity control
  •  Chamber size: 24 inches wide by 28 inches long by 15 inches high

In addition to make-up air and water vapor, three or four corrosive gases are typically used. These pre-diluted gases are fed through mass flow meters into amixing chamber prior to introduction into the MFG exposure chamber. The concentration of each gas can be controlled to the part per billion levels usingthese mass flow meters. The gases used in the test are chlorine (Cl2), hydrogen sulfide (H2S), sulfur dioxide (SO2), and nitrogen dioxide (NO2).

All of the gases have natural and man-made sources. Different gaseous pollutants dominate in different field environments. The concentrations of the corrosivegas species can range up to 2 parts per million (ppm) each, or 2000 parts per billion. The relative humidity of the gas mixture is usually maintained between70 – 75%, sometimes as high as 95%, and the temperature between 25 – 40° C (77 – 104° F). The duration of the test is variable, but is usually a maximum of 504=]hours or 21 days.

General standards:

  • ASTM B 827-05(2014) – Standard Practice for Conducting Mixed Flowing Gas (MFG) Environmental Tests.
  • ASTM B 845-97(2013)e2 – Standard Guide for Mixed Flowing Gas (MFG) Tests foe Electrical Contacts
  • ANSI/EIA-364-65A:1998, TP-65A – Mixed Flowing Gas
  • IEC 60068-2-60:2015 – Test Ke: Flowing Mixed Gas Corrosion Test
  • ISO 21207:2015 — Corrosion tests in artificial atmospheres – Accelerated corrosion tests involving alternate exposure to corrosion-promoting gases, neutral salt-spray and drying.

Industry-specific standards:

  • General Motors GMW3431.  General Procedures for Testing Switches, Section 4.4.7.
  •   Daimler Chrysler DC-10611 Rev-A, E/E Component Environment Test Specification.
  •   Comcast Drop Passive Evaluation – Splitters, Section 4.4.

Since the MFG is an accelerated testing method, the determination of acceleration factor would be helpful to understand the durability or reliabilityof the device under test. In other words, if samples can survive a certain number of days in the testing chamber, it would be very useful to approximatelyestimate how many years it can last without corrosion problems in the field.

A correlation has been drawn between accelerated test exposed and long-term field exposed samples and components in the MFG test methods developed by Battelle.  This correlation provides an acceleration of 2 days in the MFG chamber for 1 year in the field.  Thus, a 14- day exposure test would be equivalent to 7 years in service.

Contact Matergenics Pittsburgh team to discuss your needs.  We will consider all aspects of your problem and will respond promptly with a proposal for your work, including methods, costs and technical solutions. Please call Dr. George Bayer at 412-788-2163 or Dr. Zee at 412-952-9441 if you have any questions or concerns.